南美超级胎牛血清999尝鲜特惠转染试剂低至5折基质胶阶梯特惠积分商城,好礼送不停南美超级胎牛血清999尝鲜特惠转染试剂低至5折基质胶阶梯特惠积分商城,好礼送不停

Rapid test strip for HbA1c diagnosis based on fluorescent affinity immunochromatography

A rapid test strip for diagnosing glycosylated hemoglobin (HbA1c) based on fluorescent affinity immunochromatography

Chaoman Ang, Doudou Lou, Linling Hu, Wei Chen, Yefei Zhu, Zhirui Guo, Ning Gu, Yu Zhang

Journal:ANALYTICAL SCIENCES

IF:1.36

DOI:10.2116/analsci.18P135

PMID:29863029

Published:2018-06-01

research field:医学诊断分析化学结构生物学环境生物学遗传学与基因组学生物技术

Abstract

In this research, we developed a rapid and easy-to-operate point-of-care testing (POCT) strip based on fluorescent affinity immunochromatography to quantitatively determine HbA1c concentrations in whole blood. This assay, based on a sandwich method performed on test strips, effectively utilized the principle of an affinity chromatography column, which was commonly used in the detection of HbA1c, and the technology of traditional fluorescence immunochromatographic test strips (FICTS) were combined. In our test strips, the test line of traditional FICTS was transformed into the region of affinity chromatography, while improving the linearity and reducing the interference of the precursor of HbA1c and hemoglobin variants. The test strips could quantitatively detect HbA1c over a wide range (3–13.8%) with excellent linearity ( R 2 > 0.99), and the assay accuracy was demonstrated by comparing with high-performance liquid chromatography (HPLC) ( R 2 > 0.95). The simple, rapid, effective and quantitative strips will provide a novel method for the detection of HbA1c in clinical diagnosis.